• Wiring DiagramPage 3-2• Programmingthe DisplayPage 5-13• CalibrationChapter 6• TroubleshootingChapter 7QUICK GUIDEInstruction Manual19 Thomas, Irvin

Strany 2 - Table of Contents

Panel mountingPanel cut-out: 138 mm x 138 mm (5.4” x 5.4”)Front panel dimensions: 144 mm x 144 mm (5.7” x 5.7”)Included hardware: 2 flat head screws

Strany 3

Model 14 Oxygen AnalyzerVertical pipe mounting Included hardware: 2 pan head screws, 4 mm x 60 mm long for red clamping bow Horizontal pipe mounting

Strany 4 - Introduction

Model 14 Oxygen AnalyzerChapter3: Installation of the Instrument3.1 UnpackingThe analyzer should be unpacked with great care.Watch for any loose acces

Strany 5

Model 14 Oxygen Analyzer3.8 Electrical Connections Model 14Do not switch the instrument on until completion of the installation.An aluminum armor pla

Strany 6

Model 14 Oxygen Analyzer3.9 Terminal Descriptions3-3DESCRIPTION CONNECTION0 - 20 mA or 4 - 20 mA (n°1) [+] user 0 - 20 mA or 4 - 20 mA (n°1) [-] use

Strany 7

Model 14 Oxygen Analyzer3-43.10 Changing the Programming LanguageThe default programming language is English. To change the language, follow the proce

Strany 8 - Description of the Analyzer

Model 14 Oxygen Analyzer3-53.11 Programming the Power Supply FrequencyThe power supply frequency can be changed if necessary. This change occurs at th

Strany 9

Model 14 Oxygen Analyzer3-6

Strany 10 - Model 14 Oxygen Analyzer

Model 14 Oxygen AnalyzerChapter4: Using the Instrument4.1 Front Panel KeysThe display may be programmed to indicate : Sample concentration Sample te

Strany 11

Model 14 Oxygen Analyzer4-24.2 Displays 1 to 4 (live displays)100% : Dissolved Oxygen measurement 23.2°C : Temperature measurementO2 : ApplicationS1..

Strany 12 - Chapter3:

Model 14 Oxygen AnalyzerTable of ContentsiChapter 1 : Introduction ...

Strany 13

Model 14 Oxygen Analyzer4.3 Description of the Function KeysThe function keys below will be highlighted at the bottom of the screen :4-34.4 IconsWait

Strany 14

Model 14 Oxygen Analyzer4.5 Enter or Modify a ValueThe highlighted digit can be modified with the key .Each digit can be validated by pressin

Strany 15 - 23.2°CO2

Model 14 Oxygen AnalyzerChapter5: Programming the Transmitter5.1 Main Menu5-110023.2°CO2% SatDisp2 MenuS1 S3S2 S4CONC. CALIB.TEMP. CALIB.PARAMETERSCAL

Strany 16

Model 14 Oxygen Analyzer5.1.1 CALIBRATION Menus➯ Any calibration should follow the procedure below:1. Configure the calibration characteristics in the

Strany 17

Model 14 Oxygen Analyzer5.1.2 MAINTENANCE Menu5-321.6°C60.5 nA21.0 ppmMAINTENANCEAn access code may be required.See page 5-14 for CODE menu.mUsed for

Strany 18 - Using the Instrument

Model 14 Oxygen Analyzer5-4TEMP. COMP.MEASURESelectSelectP: 0760.0 mm HgTEMP. COMP.SALINITY COMP.TYPE: ManualTEMP.: 25.02˚C EnterPROGRAMMINGSelectMEAS

Strany 19

Model 14 Oxygen Analyzer5-55.1.3.2 ALARMS MenuRelays S1 through S4 may be allocated to the limit, alarm system or timer functions.ALARMSSelectALARM 1

Strany 20

Model 14 Oxygen Analyzer5-6➯ SYSTEM ALARM FUNCTION:Relays S3 and S4 may be used as a fault indicater. To control the faults traced by the analyzer, co

Strany 21

Model 14 Oxygen Analyzer5-7➯ TIMER FUNCTION:Relays S3 and S4 may be programmed to a timer function.ALARM 4SelectMODE: TimerINTERV: 1440mmIMPUL. Nb.: 5

Strany 22 - Programming the Transmitter

Model 14 Oxygen Analyzer5-85.1.3.3 mA OUTPUTS MenuThe analog output signals allow the transmission of the measurements from the analyzer to any extern

Strany 23

Model 14 Oxygen AnalyzerTable of ContentsChapter 5 : Programming the Analyzer (Continued) ...

Strany 24 - 21.0 ppm

Model 14 Oxygen Analyzer5-9AFFECT - Conc. To set the choice of analog output to the- nA concentration or temperature measurement- ˚C/˚FTYPE - 0/20 C

Strany 25 - PROGRAMMING

Model 14 Oxygen Analyzer5-10N° XX Enter number (0 – 32) BAUD - 300 Transmission speed in baud- 600- 1200- 2400- 4800- 9600 PARITY - No Without p

Strany 26

Model 14 Oxygen Analyzer5-115.1.4 SERVICE MenuCALIBRATIONAVERAGESOFT ISSUESelectSelectAVERAGE: 9TESTDISPLAYSelectCONC: ppmTEMP: ˚CPRESSURE: mmHgLANGUA

Strany 27

5.1.4.1 AVERAGE MenuAn access code may be required.See page 5-14 for CODE menu.mProgram a moving average on the concentration measurement.The measurem

Strany 28 - Timer Function

5.1.4.2 DISPLAY MenuDISPLAYCONC: ppmTEMP: ˚CPRESSURE: mmHgLANGUAGE: GBEnterSERVICESelectAVERAGEDISPLAYCODESOFT ISSUEDEFAULT VAL.ADJUST mACONFIGURATION

Strany 29 - SPECIAL PROG. OUTPUTS 1/2

5.1.4.3 CODE MenuProtection codes may be programmed for access to the PROGRAMMING, CALIBRATION, and SERVICE menus.This code is deactivated by entering

Strany 30 - Special Prog

5.1.4.4 SOFT ISSUE MenuThis menu displays the software version installed in the instrument.LoadingDEFAULT VAL.YesmValues?DefaultEnterSERVICESelectAVER

Strany 31

5.1.4.6 mA ADJUST MenuThe analog output signals are factory-adjusted (upper limit: 20mA). However, if the upper limit of one of the outputs, drifts, t

Strany 32 - Model 1400

Model 14 Oxygen AnalyzerChapter 6:Calibrating the Instrument6.1. Calibrating the Temperature SensorThe temperature sensor is located in the dissolved

Strany 33

Model 14 Oxygen Analyzer6.1.1. Automatic Temperature CompensationThe sensor continuously measures the sample temperature. The concentration values are

Strany 34 - LANGUAGE: GB

Model 14 Oxygen AnalyzerChapter1:Introduction1.1 IntroductionThe Model 14 Dissolved Oxygen Analyzer is a single-channel analyzer for the measurement o

Strany 35 - SERVICE: 0000

Model 14 Oxygen Analyzer6.1.2. Manual Temperature Compensation Manual temperature compensation should be used only when the sample temperature is cons

Strany 36

Model 14 Oxygen Analyzer6.2. Measurement Calibration The slope calibration can be done in air or water. The zero can be set with an electronic zero or

Strany 37 - CONFIGURATION CONFIGURATION

Model 14 Oxygen Analyzer6.2.2. Slope Calibration in Air with a Chemical Zero6-5ZERO CALIB.CAL21.2ppm25.0 ˚COKEXECUTIONSelectPROGRAMMINGCAL0ppm25.0 ˚CO

Strany 38 - Calibrating the Instrument

Model 14 Oxygen Analyzer6.2.3. Slope Calibration in Process with an Electronic Zero6-6CONC. CALIB.ZERO CALIB.ZERO CALIB. SLOPE CALIB.SLOPE CALIB.Selec

Strany 39 - 21.2 21.2

Model 14 Oxygen AnalyzerSLOPE :The flashing “CAL” message indicates that theinstrument is in calibration mode. Wait for the current to stabilize(appro

Strany 40

Model 14 Oxygen Analyzer6-8

Strany 41

Model 14 Oxygen AnalyzerChapter 7:Start-up and Troubleshooting7.1 Start-up7.1.1 Probe ConnectionSee page 3-3 for the terminal connection. 7.1.2 Main

Strany 42

Model 14 Oxygen AnalyzerCauses and solutions:A) The membrane permeability has changed (coating).➯ Calibrate the analyzer and check if the concentratio

Strany 43

Model 14 Oxygen Analyzer7.3 Troubleshooting the Electronics7-3WARNING!Never attempt to service before disconnecting the instrument from the main power

Strany 44

Model 14 Oxygen Analyzer7-4Interferences Are there any sources of potential interference, chemical, external, temperature, pressure, etc.?Cable shield

Strany 45

Model 14 Oxygen Analyzer1.3 Main Characteristics Range: 0 - 299.99% saturation; 0 - 29.99 ppm Calibration in the air Temperature compensation Pr

Strany 46 - Start-up and Troubleshooting

Model 14 Oxygen AnalyzerChapter 8: Error MessagesDuring failure, the measurement is replaced by dashes “ - - -.”8-1CONCENTRATIONTOO HIGH- - -STAND-BY1

Strany 47

Model 14 Oxygen Analyzer8-2Error messages during calibrationThe current value exceeds the limits. Confirmthere is no short-circuit in the measuring li

Strany 48

Model 14 Oxygen AnalyzerAppendix 1: Pressure Conversion TableA-1hPa or mbars mm Hg mm H2O psi inches Hg inches H2O 920.00 690 9503.60 13.34 27.1

Strany 49

Model 14 Oxygen AnalyzerAppendix 2: Temperature Conversion TableA-2• Conversion from ˚C into ˚F : ˚F = 1.8 *˚C + 32• Conversion from ˚C into ˚K : ˚K =

Strany 50 - Chapter 8: Error Messages

Model 14 Oxygen AnalyzerAppendix 3: Default ValuesCONC. CALIB. PARAMETERSZERO CALIB. DATE :01/01/98Type : Elec. Auto ZERO: 0.00 nASLOPE S: 3.000 nA/pp

Strany 51

Model 14 Oxygen AnalyzerA-4AVERAGEAVERAGE : 1 DISPLAYDISPLAYCONC. : ppb/ppmTEMP. : °CPRESSURE : mmHgLANGUAGE : GB CODECODECALIB. : 0000PROG. : 0000SE

Strany 52 - Pressure Conversion Table

Model 14 Oxygen AnalyzerA-5Broadley Technologies Corporation19 ThomasIrvine, California 92618 USACorporate Headquarters:Toll-Free: 877.246.7900 Tel:

Strany 53 - Temperature Conversion Table

Model 14 Oxygen AnalyzerDisplay Display in concentration units or % saturationDisplay of the direct cell current in nADisplay of the sample temperatur

Strany 54 - Appendix 3: Default Values

Model 14 Oxygen Analyzer1.5 Dimensions(Dimensions are in mm and inches)1-4Esc Enter33.5 mm24 mm 0.936"1.319"24 mm 0.945" 28 mm 1.102&qu

Strany 55

Model 14 Oxygen AnalyzerChapter2: Description of the Analyzer2.1 Transmitter2.1.1 Presentation of the Transmitter2-1The electronic unit amplifies the

Strany 56

Model 14 Oxygen Analyzer2-22.1.2 Transmitter Schematic:1 : Programmable potentiostat 6 : Auxiliary input 2 : Polarization voltage amplifier 7 : Mul